Installation requirements

LOCATION

The location, room number and exact position of the BSCs should be determined and communicated prior to installation. We emphasize that the BSCs should not be placed near doors, open windows, under or near exhaust vents, in order to achieve optimal safety. Busy “walkways” between the laboratory tables and the lab layout, where there is a lot of walking around, should be avoided at all times. Please consult your Health, Safety and Environment Officer and/or BioSafety Officer.

CONNECTING TO AN EXHAUST DUCT SYSTEM

The connection of the BioSafety Cabinets (BSCs) / Microbiological Safety Cabinets (MVKs) to the return air duct is not part of our delivery and services as standard. This work must be carried out by third parties. Labolutions can only validate the BSCs / MVKs if they are connected to the return air duct if required. Usually the BSC’s are class II (in the past also called IIa and IIb) recirculation principle and the exhaust passes a HEPA filter before it is blown into the (lab) space. However, the BSCs can also be connected to the exhaust system by means of a direct and/or thimble connection. The supply of the direct and/or thimble unit as well as the connection of it is never included in our offer(s). Labolutions also assumes that the work can be performed in succession by our colleagues and / or subcontractors. Of course, it can be discussed in the preliminary process, to connect these BSCs / MVKs by Labolutions. This is definitely not our preference! It is customary to let your (home-) installer do this work.

CONNECTING TO AN EXHAUST DUCT SYSTEM

The connection of the BioSafety Cabinets (BSCs) / Microbiological Safety Cabinets (MVKs) to the return air duct is not part of our delivery and services as standard. This work must be carried out by third parties. Labolutions can only validate the BSCs / MVKs if they are connected to the return air duct if required. Usually the BSC’s are class II (in the past also called IIa and IIb) recirculation principle and the exhaust passes a HEPA filter before it is blown into the (lab) space. However, the BSCs can also be connected to the exhaust system by means of a direct and/or thimble connection. The supply of the direct and/or thimble unit as well as the connection of it is never included in our offer(s). Labolutions also assumes that the work can be performed in succession by our colleagues and / or subcontractors. Of course, it can be discussed in the preliminary process, to connect these BSCs / MVKs by Labolutions. This is definitely not our preference! It is customary to let your (home-) installer do this work.

BY DEFAULT NO MEDICAL GASES AND APPLICATIONS INTENDED FOR ADMINISTRATION TO PATIENTS.

The pipes of our taps are not suitable for medical gases and medical applications, which are intended for administration to patients, as stated in the ISO 7396-1 (eg CO2, O2, N2 etc.). Please ask explicitly if connections are required for the use of medical gases or medical applications, so that we can make you an additional offer.

SPACE ABOVE BSC TO (LOWERED) CEILING

Engineering and installation advice for recirculating installation or connection to the exhaust air duct of your BioSafety Cabinets.
If the workbench is installed in a recirculating mode, we recommend keeping a free space of at least 25 cm above the exhaust HEPA filter. This free space is necessary in order to guarantee optimal personal protection!
If there is not enough free space above the “HEPA filter outlet” on the upper side of the BSC, a counter pressure can occur!
If there is not enough free space towards the ceiling, above the “HEPA filter outlet” on the top of the BSC, a counter pressure can arise which could possibly compromise the operator protection. This space is also needed during validation work to scan the filter correctly.

RISK ANALYSIS

We advise your BioSafety Officer to make a risk analysis. If this risk analysis shows that the safety workbench(s) must be connected to an exhaust air duct (Ø250mm), we strongly advise you to have this done by means of a thimble system.
This is as advised in the EN12469 and in accordance with the NSF49 standards. Labolutions has various thimble systems from ninoSAFE in its portfolio and can supply these if required. In this way the protection of persons remains guaranteed in case of a malfunctioning drain system. An additional advantage
An additional advantage of using a thimble is that the pressure in the laboratory is not affected by switching the BSC / MVK on / off.

SUFFICIENT CAPACITY

You must at all times ensure that your drainage system has sufficient capacity to be able to set the channel, when the safety workbench is in operation, to a slight vacuum.
When the safety workbench is in operation, it must be possible to set it to a slight underpressure. When using a thimble we advise you to always remove 10-15% extra air. It is also necessary to ensure the correct position of the workbench for the connection to the exhaust duct in relation to the exhaust hood and
thimble.

EXACT PLACE DETERMINATION

Location of the MVK / BSC When placing the BioSafety Cabinets (BSCs) / Microbiological Safety Cabinets (MVKs), you should also SVP take into account the correct location of the MVK / BSC within the available space. Tip 1) Make sure that there are no air outlet or exhaust grilles of the air treatment systems just above the front of the BSC / MVK in the ceiling, this can negatively affect the inflow and outflow of the BSC / MVK and thus the biosafety Tip 2) do not place the MVK / BSC close to doors and doorways and / or location where there is often movement of colleagues Tip 3) not close to open windows (think of proper lab design advice)

VALIDATION

Standard validation or execution IQ-OQ Installation Qualification – Operational Qualificatio. The BioSafety cabinets are validated with our filter efficiency test as standard after installation. This according to the WB 2.2. protocol and for workbenches where GMP speeds have been chosen by means of WB 2.2.2. For pharmacies and pharmaceutical companies there are often strict (GMP) rules that must be adhered to. IQ-OQ is a part of this. If ordered we can carry out the IQ-OQ at your location.

KI TEST

KI tests KI (potassium iodide) disc tests are not included in our standard activities. We can optionally combine additional KI testing with our services and have it performed in accordance with the requirements of EN12469/DIN 12980. If there are any cooperated and/or company protocols, we kindly ask you to indicate these at the time of commissioning. The performance of the KI test must be communicated prior to the planning of the work (order provision) included. This in connection with the planning and implementation of the necessary equipment.

HYGIENIC HERMETIC FINISH

Hermetic hygiene finish of the BSC adjoining the walls/walls – A hermetic smooth finish of the BioSafety cabinets to walls and/or ceilings is often desired and/or required in microbiological laboratories and R& D facilities, pharmacy areas and/or clean rooms. In our quotations, a smooth finish from the supplied workbench(s) to the wall and/or ceiling is not included as a standard feature, unless this has been explicitly requested in the preliminary process and/or in the URS (User Requirement Specifications) and thus included in our quotation. We kindly request you to state this clearly at the start of the project in order to avoid unnecessary delays in the installation and validation.